SKC: Shaping the Future of Air Sampling  

 

 

EVM-7 Environmental Monitor

2-in-1 Particulate and Gas Concentration Measurement

EVM-7 Key Features:

  • Simultaneous particulate mass and gas concentration measurement

  • Twist and click impactor for PM2.5, PM4, PM10, or TSP* measurement without disassembly and external cyclones

  • 90-degree light scattering photometer for mass particle concentration measurement and real-time display

  • Built-in compensating sample pump for simultaneous gravimetric sampling on a 25 or 37-mm filter cassette inside the instrument.

  • Smart auto-recognized toxic gas sensors

  • High-performance 10.6 eV PID for VOCs


  • * Within the instrument’s measurement range

EVM-7 Environmental Monitor 755-EVM7
EVM-7 Environmental Monitor

 

EVM Monitor Spec. Sheet (PDF)

Monitor multiple air quality hazards/parameters using one instrument with the Quest® EVM-7 environmental monitor. EVM-7 simultaneously measures particulates, volatile organic compounds (VOCs), toxic gas, carbon dioxide, relative humidity, temperature, and air velocity (with accessory probe). Two instruments in one means lower cost of ownership and elimination of multiple instruments.

Additional EVM-7 Features:

  • Time history datalogging

  • Factory calibrated relative to ISO 12103-1, A2 (Arizona Road Dust)

  • Site correction factors for particulates (user supplied)

  • Auto run and security code protection for unattended monitoring

  • Carbon dioxide trending algorithms for calculation of room air exchange rates (with QuestSuite Pro II Software)

  • Real-time measurement with once-per-second update rate

  • Trend graphing for all parameters

  • Menu-driven user interface with 10 pushbuttons/4 softkeys

  • User selectable digital output for external functions

  • Displayed measurements of Level, Minimum, Maximum, Average, STEL, and TWA

  • Displays in 6 languages
EVM Open showing Filter Cassette
EVM Sensor Bar Open
Easy Access to gravimetric
sample filter cassette
Removable sensor bar for
easy access to sensors

EVM-7 Applications

  • Worksite monitoring
  • Industrial hygiene surveys
  • Military site evaluation
  • Aircraft and train environment sampling
  • Construction/mining site monitoring
  • Engineering studies
  • Industrial fenceline monitoring
  • Remediation sampling
  • IAQ walk-through and stationary studies
  • Air quality assessment
  • VOC screening and monitoring
  • Regulatory compliance

 

Additional EVM Models

EVM-4 and EVM-3 feature the durability, quality, and overall design of the EVM-7.

EVM-4 Comprehensive IAQ Analysis
Monitors and logs:

  • Carbon monoxide
  • Relative humidity
  • Temperature
  • Carbon dioxide (room air exchange rates calculated in QuestSuite Pro II Software)
  • Optional sensor accessories: AirProbe-10 velocity probe, PID, particulates, replace CO sensor with one of 8 other toxic gas sensors


EVM-3 Real-time Direct-reading Particulate Monitor
Measures and logs:
  • Particulate mass concentration (PM2.5, PM4, PM10, or TSP*)
  • Temperature
  • Relative humidity
  • Optional sensor accessories: AirProbe-10 velocity probe, CO or select from 8 other toxic gas sensors, CO2, PID

Performance Profile

Method
Base Units
Display Res.
Display Range
Accuracy/Repeatability
Particulates
Light-scattering Photometer
mg/m3
0.001
0.000-199.9
± 15% (rel ARD)
Particulates Size Range
µm
N/A
0.1-10
**
VOCs
10.6 eV PID
ppm
0.01
0.00-2000
± 5%/2% (rel to Isobutylene) at calibration level
Carbon Dioxide
NDIR
ppm
1
0-20,000
± 2% of signal, ± 50 ppm between 0-2500
Toxic Gas (Electrochemical Sensors)
Carbon Monoxide
ppm
1
0-1000
± 5%/2% of signal
Chlorine
ppm
0.1
0.0-20
± 5%/2% of signal
Ethylene Oxide
ppm
0.1
0.0-20
± 5%/2% of signal
Hydrogen Cyanide
ppm
0.1
0.0-50
± 5%/2% of signal
Hydrogen Sulfide
ppm
1
0-500
± 5%/2% of signal
Nitric Oxide
ppm
0.1
0.0-100
± 5%/2% of signal
Nitrogen Dioxide
ppm
0.1
0.0-50
± 5%/2% of signal
Oxygen
%
0.1
0.0-30
± 5%/2% of signal
Sulfur Dioxide
ppm
0.1
0.0-50
± 5%/2% of signal
Temperature
Junction Diode
deg C
0.1
-10.0- 60.0
± 1.1 deg C
deg F
0.1
14.0- 140.0
± 2 deg F
Relative Humidity
Capacitive
% humidity
0.1
0.0-100
± 5% RH of signal between 10 - 90%
Air Velocity
Omni-directional Heated Thermistor Windprobe
m/sec
0.1
0.0-20
± 0.12 m/s + 4.5% of signal
ft/min
1
0-3940
± 23.6 ft/min + 4.5% of signal
General
Size Fractions PM2.5, PM4, PM10, or TSP**
Sample Pump Flow Rate 1.7 L/min
Power/Run Time Rechargeable Li-Ion battery pack provides minimum of 8 hrs of continuous operation
Weight 2.9 lb (1.3 kg)
Standards CE marked and RoHS compliant
Arizona Road Dust (ARD)
** The photometer can detect particles up to 100 µm; however, accuracy is reduced for sizes greater than 10 µm.

Ordering information

EVM Monitors
Cat. No.
EVM-7 IAQ/Dust Monitor with 5 Sensors: VOCs, CO2, Dust, Temperature, and Humidity includes datalogging monitor with sensors for VOCs (PID), CO2, particulates, temperature, and humidity (with air probe input connector), two 37-mm filter cassettes (for pump protection), 12-V DC universal power supply, USB cable, silicone dielectric grease, PID cleaning kit, 10-pk pump protection filters, dust zero calibration assembly, toxic gas calibration adapter, and storage case. Purchase Air Probe-10 Air Velocity Probe separately.
EVM-7 CO/IAQ/Dust Monitor with 6 Sensors: VOCs, CO2, CO, Dust, Temperature, and Humidity includes datalogging monitor with sensors for VOCs (PID), CO2, CO, particulates, temperature, and humidity (with air probe input connector), two 37-mm filter cassettes (for pump protection), 12-V DC universal power supply, USB cable, silicone dielectric grease, PID cleaning kit, 10-pk pump protection filters, dust zero calibration assembly, toxic gas calibration adapter, and storage case. Purchase Air Probe-10 Air Velocity Probe separately.
EVM-4 IAQ Monitor with 3 Sensors: CO2, Temperature, and Humidity includes datalogging monitor with sensors for CO2, temperature, and humidity (with air probe input connector), 12-V DC universal power supply, toxic gas calibration adapter, USB cable, and storage case. Purchase Air Probe-10 Air Velocity Probe separately.
EVM-4 CO/IAQ Monitor with 4 Sensors: CO2, CO, Temperature, and Humidity includes datalogging monitor with sensors for CO2, CO, temperature, and humidity (with air probe input connector), 12-V DC universal power supply, toxic gas calibration adapter, USB cable, and storage case. Purchase Air Probe-10 Air Velocity Probe separately.
EVM-3 Dust Monitor with 3 Sensors: Dust, Temperature, and Humidity includes datalogging monitor with sensors for particulates, temperature, and humidity (with air probe input connector), two 37-mm filter cassettes (for pump protection), 12-V DC universal power supply, USB cable, silicone dielectric grease, 10-pk pump protection filters, dust zero calibration assembly, and storage case. Purchase Air Probe-10 Air Velocity Probe separately.
Dummy Sensor, seals empty sensor socket
Carbon Monoxide Smart Sensor
Chlorine Smart Sensor
Ethylene Oxide Smart Sensor
Hydrogen Cyanide Smart Sensor
Hydrogen Sulfide Smart Sensor
Nitric Oxide Smart Sensor
Nitrogen Dioxide Smart Sensor
Oxygen Smart Sensor
Sulfur Dioxide Smart Sensor
AirProbe-10 Air Velocity Kit includes AirProbe-10, mounting bracket, adapter cable, AC charger, and carry case
120 V
220 V

Additional accessories are available. Contact SKC

 

EVM Monitor Spec. Sheet (PDF)