High flows from 5 to 15 L/min
Achieve flow rates that you couldn't reach before when using a personal sample
pump. The Leland Legacy® provides the high flows and long run times
of a vacuum-style pump in a compact, portable, and battery-operated sampler.
Long runs for 24 hours on one battery charge
Rechargeable lithium-ion (Li-Ion) battery pack provides 24-hour run times
with impactors and other sampling devices with low back pressures*.
Leland Legacy is ideal for 24-hour indoor air studies and unattended ambient
air sampling.
| * |
Leland Legacy is not recommended for high back pressure
applications such as asbestos clearance sampling. |
Typical Run Time†
24+ hours:
- Sioutas Impactor at 9 L/min (approximately 13 inches water
back pressure)
- Personal Environmental Monitor (PEM) with 37-mm, 2.0-µm
PTFE filter 225-1709‡ at
10 L/min
- Low-volume PUF Tube at 5 L/min
- IMPACT Sampler (PM2.5 or PM10) at 10 L/min
- For extended run times, the pump may be operated while attached
to the approved charger.
| † |
Results obtained using a new pump and new fully charged
battery. Pump performance may vary. |
| ‡ |
Back pressure on PTFE filters can vary within
the same lot. Maximum operating temperature is 464 F (240
C)
based
on support ring. |
Low Levels
Longer runs at higher flows provide more sample for the enhanced
sensitivity required for measuring low concentrations, particularly
in indoor air environments.
Low Noise
With the noise-reducing case accessory, the Leland Legacy can
operate at 10 L/min and 12 inches water back pressure with
a noise level
of only 52 dBA.**
** Measured
3 feet (1 meter) from pump operating at 10
L/min without media
Advanced flow control
The patented (U.S. Patent No. 5,892,160) internal flow sensor measures
flow directly and acts as a secondary standard to constantly maintain
the set flow. Set flow is achieved immediately at start-up
and flow calibration is automatically maintained
by built-in sensors that compensate for differences in temperature
and atmospheric pressure during sampling.
Flexible programming
- Manual three-button operation
No tools needed. Simply use the built-in keypad to set, calibrate, and
sample! Real-time sampling parameters are available at the touch of a
button.
- PC programmability
The ultimate in pump scheduling and record keeping. Use Leland Legacy with
your PC and DataTrac for Leland Legacy Software to create complete running
sequences, download sampling and calibration data, and generate sampling
reports for ISO 9000, TQM programs, and occupational health and safety
management systems. DataTrac Software for Leland Legacy is included in
the 5-pack
kits and sold separately.


Accurate automatic calibration
The patented (U.S. Patent
No's. 6,227,031 and
6,363,769) built-in CalChek feature
provides direct communication to a Defender primary standard calibrator
for fast and easy calibration without manual adjustments. The Defender
calibrator
is
sold separately
or available in convenient Leland Legacy CalChek pump kits.
Leland Legacy Sampler Applications
Use in non-explosive locations only.
- 24-hour exposure monitoring
- Indoor air studies
- Particulate sampling:
- Pesticides — EPA TO-10A and ASTM 4861
- Bioaerosols
(spore traps/cassettes)
- Cotton dust
- Radon decay products (ASTM D6327-98)
- Isokinetic sampling
- Environmental sampling
- Sampling with cascade impactors
- Spore trap sampling (VersaTrap)
- PM2.5 with PEM at 10 L/min (EPA IP-10A)
The Leland Legacy Advantage
Collects more sample
High flows and long run times greatly increase the amount of
particulate matter collected when sampling low levels.
This
enhances the usefulness of the data for environmental health and epidemiology
studies.
Designed for indoor
environments
Noise, size, and weight are minimized relative to the high flow
range.
Downloadable history
Critical sampling parameters are retained in
memory during the sampling period and can be downloaded for record keeping.
Leland Legacy and Sioutas Impactor* 
Suitable for use with the Leland Legacy Pump at 9 L/min, the Sioutas Impactor
separates airborne particles precisely into five size ranges from ultrafine
to PM2.5. When used with PTFE filters, the Sioutas Impactor is highly
efficient at collecting particles without using impaction grease or coated
substrate and at retaining unstable compounds for size-fractionated chemical
analysis. The performance of the Leland Legacy pump with the Sioutas Impactor
has been verified through EPA-ETV. Click
here for more information on the Sioutas Impactor.
Click
here for the SKC Update to EPA IP-10A (PDF)
* Developed by Dr. Constantinos Sioutas of
the University of Southern California in partnership with NUATRC.
Leland Legacy
and Sioutas Impactor Receive ETV Verification
U.S. EPA Environmental Technology Verification (EPA-ETV) is a program that
furthers environmental protection by accelerating acceptance and use of improved,
cost-effective technologies through performance verification. EPA-ETV recently
tested the SKC Leland Legacy Sample Pump with the Sioutas Impactor and released
its report. For a verification report on the Leland Legacy and Sioutas performance,
visit the AMS Center at www.epa.gov/nrmrl/std/etv/vtams.html.
The SKC Leland Legacy and NUATRC
The Leland Legacy was developed in partnership
with the Mickey Leland National Urban Air Toxics Research Center (NUATRC) to
further advance personal exposure
measurement.
Established by Congress in the Clean Air Act Amendments of 1990 (CAAA), the non-profit
Leland Center is responsible for investigating public health risks posed by air
toxics to urban populations. The Leland Center is a private/public research partnership
that brings together corporate, academic, and government scientists to develop
a sound peer-reviewed air toxics research program. This program is jointly supported
by the private sector and the U.S. EPA.
Performance Profile
Leland Legacy Sample Pump
| Flow Range |
5 to 15 L/min |
| Flow Control System |
Closed loop with patented internal
flow sensor
U.S. Patent No. 5,892,160 |
| Compensation Range |
15 L/min at 5 inches water back pressure
10 L/min at 12 inches
water back pressure
5 L/min at 20 inches water back pressure |
| Accuracy |
Flow rate: ± 5% of set-point after calibration
to desired flow |
| Accuracy |
Timing: 1 min/mo at 25 C |
| Accuracy |
Atmospheric Pressure: ± 0.3 in Hg |
| Flow Fault |
If the pump is unable to compensate for > 15 seconds due
to excessive back pressure, a flow fault icon displays and flashes,
the pump enters Hold mode, and the pump retains historical data.
Auto restart is attempted every 20 seconds up to 10 times. Adjustable
with DataTrac for Leland Legacy software |
| Display |
LCD, displays pump serial number, pump software revision level,
flow rate, volume, temperature, atmospheric pressure, time of
day, run time, and pump status, i.e., hold and run as well as
Setup information |
| Volume Display |
Continually updated, based on corrected flow rate multiplied
by sampling time |
| Time Display |
Time of day in hours and minutes (12 or 24-hour clock) with
AM and PM indicators |
| Timer Display Range |
1 to 99999 minutes (69 days). If the run time exceeds 69 days,
the timer display rolls over. |
| Operating Temperature Range |
32 to 113 F (0 to 45 C)
Protect sample pump from weather when in use outdoors. |
| Typical Run Time† |
24+ hours:
- Sioutas Impactor at 9 L/min (approximately 13 inches
water back pressure)
- Personal Environmental Monitor (PEM) with 37-mm, 2.0-µm
PTFE filter 225-1709‡ at
10 L/min
- Low-volume PUF Tube at 5 L/min
- DPS Sampler (PM2.5 or PM10) at 10 L/min
- For extended run times, the pump may be operated while
attached to the approved charger.
| † |
Results obtained using a new pump and new fully
charged battery. Pump performance may vary. |
| ‡ |
Back pressure on PTFE filters can vary within
the same lot. Maximum operating temperature is 464 F
(240 C) based on support ring. |
|
| Noise Level |
62.5 dBA — pump without case
52 dBA — pump housed in noise-reducing
case (optional accessory Cat. No. 224-89) measured 3 ft (1
m) from pump operating at 10 L/min without media |
| User-adjustable Values |
Sample run time, calibration, clock display, flow rate, time
of day, delayed start, and temperature and atmospheric pressure
display |
| Recorded Values |
Start date and time, stop date and time, total sample time,
flow rate, sample volume, temperature, atmospheric pressure,
and pump mode transitions |
| Adjustable Logging Interval |
Records pump history from 3 seconds (15.4 min of data) up
to 8 hours (over 102 days of data) depending on setting. Option
available when using DataTrac Software |
| Power Supply |
Battery: removable, rechargeable lithium-ion (Li-Ion) battery
pack 7.4 V x 12-Ah capacity, 88.8 Wh. (See Li-Ion Battery
Shipment below.)
Charger/AC
adapter: input voltage 100-240 V AC |
| Battery Recharge Time (with SKC-approved charger) |
15 hrs |
| Temperature: Charging |
32 to 113 F (0 to 45 C) |
| Temperature: Storage |
-4 to 95 F (-20 to 35 C) |
| Size |
8 x 3.9 x 2.6 in (20 x 10 x 7 cm) |
| Weight |
36 oz (1 kg) |
| Case |
Thermoplastic with soft rubber overmolding |
| RFI/EMI Shielding |
CE marked |
Ordering Information
Pump Kits and Accessories*
Use
in non-explosive locations only;
not UL Listed for intrinsic safety.
Leland
Legacy Pumps contain Li-Ion batteries and may be subject to special
shipping
regulations. |
Cat. No. |
Leland Legacy Pump* with
Li-Ion battery pack
Use in non-explosive
environments only; not UL Listed for intrinsic safety; CE marked
Requires
charger; see kits or accessories below |
100-3002 |
Single Pump Kit* includes
Leland Legacy pump, Li-Ion battery pack, single charger/adapter unit, and
deluxe carry case, CE marked
Use in non-explosive locations
only; not UL Listed for intrinsic safety
|
100-240 V |
100-3002K |
Single Pump CalChek Kit* includes
Leland Legacy pump, Li-Ion battery pack,
single charger/adapter unit, Defender calibrator (Revision Level
C, CalChek ready) with charger and CalChek communication cable,
and Pelican carry case. CE marked
Use in non-explosive locations
only; not UL Listed for intrinsic safety
|
100-240 V |
100-3002DFK |
| Single Pump Starter Kit includes pump with battery,
charger, Tygon tubing (3 feet, ¼-inch ID), and collar clip
with cable tie |
100-240 V |
100-3002-S |
5-pack Pump Kit* includes
5 Leland Legacy pumps, 5-station charger,
and DataTrac Software, in a Pelican case. CE marked
Use in non-explosive locations
only; not UL Listed for intrinsic safety |
100-240 V |
100-3002K5 |
5-pack CalChek Kit* includes
items in 5-pack Leland
Legacy pump kit, CalChek Kits, Li-Ion battery packs, DataTrac Software,
and Defender calibrator (Revision Level C, CalChek
ready) with charger, and CalChek communication cable, in a Pelican
case. CE marked
Use in non-explosive
locations only; not UL Listed for intrinsic safety |
100-240 V |
100-3002DFK5 |
| Pump Certificate
|
| Certificate of Compliance, optional, must
be ordered when ordering pump |
P1501 |
| Software
|
DataTrac Software for Leland Legacy includes software
CD, DataTrac adapter, and DataTrac cable
Requires Windows 98 or higher and available serial port or appropriate
adapter compatible with PC and system
Download
DataTrac Software for Leland Legacy |
877-92 |
| Calibrators and Accessories
|
Defender 510 Model, 300 to 30,000 ml/min includes
lead-acid battery, multi-plug charger (100-240 V), Optimizer Software,
39-inch (1-meter)
serial cable, and leak test caps
Software requires Windows® 2000, XP (sp 2 or 3), or
Vista (32-bit version) and RS-232 port or USB to RS-232 adapter. |
 |
|
Defender 520 Model, 300 to 30,000 ml/min includes
lead-acid battery, multi-plug charger (100-240 V), Optimizer Software,
and 39-inch
(1-meter) serial cable, and leak test caps
Software requires Windows® 2000, XP (sp 2 or 3), or
Vista (32-bit version) and RS-232 port or USB to RS-232 adapter. |
 |
|
Defender 530 Model, 300 to 30,000 ml/min includes
lead-acid battery, multi-plug charger (100-240 V), Optimizer Software,
39-inch (1-meter)
serial cable, and leak test caps
Software requires Windows 2000, XP (sp 2 or 3), or Vista (32-bit
version), and an RS-232 port or USB to RS-232 adapter. |
 |
|
CalChek Communication Cable,
provides direct communication between a CalChek-compatible AirChek 2000
or Leland Legacy pump and a CalChek-capable
Defender primary standard calibrator; included in CalChek kits above
Note: For use only with Defender calibrators
with Revision Level C or higher |
210-502 |
| Replacement Defender Lead-acid Battery Pack |
|
| Defender Universal Charger, 100-240 V, with
U.S., Euro, and Australia plugs |
|
| Defender Universal Charger, 100-240 V, with
U.S. plug |
|
| Chargers and Accessories
|
| Single Pump Charger for
XR5000 and Leland Legacy includes charging unit and wall cube
(100-240 V worldwide power input with interchangeable plugs included) |
 |
223-241 |
Take Charge 5 for XR5000 and Leland Legacy, 5-station
charger for Li-Ion batteries (100-240 V)
Use in non-explosive locations only; not UL Listed for
intrinsic safety |
 |
223-441 |
| Single Wall Cube/Power Supply, 100-240
V, with multiple plugs, included as part of the 223-241 Charging Kit |
 |
223-245 |
Li-Ion Charging Unit, 100-240 V, included
as part of the Leland Legacy and AirChek XR5000 Charging Kit, 223-241
Note: Must be used in conjunction with the 223-245
Wall Cube Charger/Power Supply
|
 |
P22300 |
| Replacement Li-Ion Battery Charging
Unit Cable, for charger 223-241 |
P75632 |
| Battery Charging Adapter, for
charging batteries outside the pump |
 |
223-248 |
| Pump Carry Cases
|
Noise-reducing Nylon Case, black, with lining to
reduce pump noise from 62.5 to 52 dB,
includes waist belt and shoulder strap
Measured 3 feet (1 meter) from pump operating at 10 L/min without
media |
 |
224-89 |
| Leland Legacy Single Case, Pelican with
foam |
224-912 |
| 5-pack, Pelican Leland Legacy Case |
224-905 |
| Replacement Parts and Tools
|
| Replacement Leland Legacy Battery Pack, Li-Ion |
P75692NUL |
| Replacement Filter/O-ring Set includes
5 filters and 1 O-ring |
P40021B |
| Replacement Inlet Filters, pk/50 |
P40021A |
| Replacement PC Board, for Leland
Legacy Sample Pump, main |
P7954802 |
| Replacement Inlet, for Leland
Legacy Pump |
P20198 |
| Replacement Belt Clip, for
Leland Legacy Pump |
P51823 |
| Leland Legacy Case includes front and back
cases; does not include serial number label |
|
| Leland Legacy Control Board |
|
| Leland Legacy Motor |
|
| Leland Legacy Inlet Side Pulsation Dampener |
|
| Replacement Yoke/Crank Arm Insert Assembly |
|
| SKC Universal Screwdriver Kit, for
use with sample pumps, includes 3 interchangeable screwdriver heads |
224-11 |
Protect sample pump from weather when in use outdoors.
| * |
Li-Ion Battery
Shipment
Rechargeable lithium-ion batteries for use with SKC sample pumps have
been tested in accordance with the UN Manual and are proven to meet requirements
of each test in the UN Manual of Test Criteria, Part III, subsection
38.3. They have a watt-hour (Wh) rating below 100.
Per 2010 IATA regulations, packaging must meet the
specifications of and contain labeling and documentation required by
IATA Packing Instructions 965, 966, and 967. See IATA
Guidance
Document: Transport of Lithium Metal and Lithium Ion Batteries, Revised
for the 2010 Regulations
|
Leland
Legacy Product Data Bulletin (PDF)
Leland
Legacy Operating Instructions (PDF)
Leland Legacy Quick
Guide (PDF)
CalChek with the
Defender Operating Instructions (PDF)
Chinese
Language Leland Legacy Operating Instructions (PDF)
Battery
Pack for Leland Legacy Sample Pumps Operating Instructions (PDF)
Take
Charge 5 Li-Ion Battery Charger Operating Instructions (PDF)
MSDS
for Lithium-Ion Rechargeable Battery (PDF)
Technical
Note: SKC Sample Pump Battery Pack Characteristics and Maintenance (PDF)
Download
DataTrac for the Leland Legacy
DataTrac
Operating Instructions (PDF)
Leland
Legacy EMC Test Certificate (PDF)
Leland
Legacy FAQs
This page was last modified July 21, 2010
|